{"id":3087,"date":"2020-06-03T07:08:44","date_gmt":"2020-06-03T07:08:44","guid":{"rendered":"https:\/\/www.bronkhorst-china.com\/?page_id=3087"},"modified":"2020-06-24T09:48:52","modified_gmt":"2020-06-24T09:48:52","slug":"principleofesflow","status":"publish","type":"page","link":"https:\/\/www.bronkhorst-china.com\/service\/technologies\/principleofesflow\/","title":{"rendered":"\u8d85\u58f0\u6ce2\u6d4b\u91cf\u539f\u7406"},"content":{"rendered":"
The operation of Bronkhorst\u00ae ES-FLOW™ flow meters is based on the propagation of ultrasound waves inside a very small, straight sensor tube, without obstructions or dead spaces. At the outer surface of the sensor tube multiple transducer discs are located which create ultrasonic sound waves by radial oscillation. Every transducer can send and receive, therefore all up- and down-stream combinations are recorded and processed. By accurately measuring the time difference between the recordings (nanosecond range) the flow velocity and speed of sound is calculated. Knowing these parameters and the exact tube cross-section, the ES-FLOW™ is able to measure liquid volume flows in the range of 4 up to 1500 ml\/min.<\/p>\n
The distinctive character of this flowmeter is that it\u2019s capable to measure the actual speed of sound, meaning that the technology is liquid independent and calibration per fluid is not necessary.<\/p>\n